i
Teledyne Analytical Instruments
OPERATING INSTRUCTIONS FOR
Model 8800A & B
Trace Moisture Analyzer
HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING
SYSTEM.
PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.
HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST
FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.
ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING. BEFORE
CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/
MANAGER.
DANGER
P/N M8800A&B
12/19/01
ECO # 01-0382
Zobrazit stránku 0
1 2 3 4 5 6 ... 71 72

Shrnutí obsahu

Strany 1 - Model 8800A & B

i Teledyne Analytical InstrumentsOPERATING INSTRUCTIONS FORModel 8800A & BTrace Moisture AnalyzerHIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MA

Strany 2

Section 2: Sensor and Sampling Techniques32. Sensor and Sampling Techniques2.1 Precautions using the sensorThe HTF™ Al2O3 sensor is designed and field

Strany 3

8800 Series Trace Moisture Analyzer Instruction Manual42.2 Sensor Technical SpecificationsType: ...Hyper Thin Film HTF™

Strany 4 - Warning Labels

Section 2: Sensor and Sampling Techniques52.3.1 In-situ InstallationIn-situ installation is recommended only for measurements where the gas pressure i

Strany 5

8800 Series Trace Moisture Analyzer Instruction Manual62.3.2 Extractive InstallationFor extractive installations we recommend our sample system ESS, w

Strany 6 - Table of Contents

Section 2: Sensor and Sampling Techniques7inside these materials.• Try to run pipes to the sensor upwards, so that contaminants tend to fall back into

Strany 7

8800 Series Trace Moisture Analyzer Instruction Manual82.5 Troubleshooting unexpected readingsIf erroneous readings are suspected on a newly acquired

Strany 8 - Section 1: Introduction

Section 3: Instrument93. Instrument3.1 Precautions using the 8800 Series Trace Moisture AnalyzerThe 8800 Series Trace Moisture Analyzer uses state-of-

Strany 9

8800 Series Trace Moisture Analyzer Instruction Manual103.2 Instrument Technical SpecificationsEnclosure: ...8800B tabletop or s

Strany 10

Section 3: Instrument113.3 Installation3.3.1 Instrument Mechanical InstallationThe 8800 Series Trace Moisture Analyzer is available in several differe

Strany 11

8800 Series Trace Moisture Analyzer Instruction Manual123.3.2 Electrical ConnectionsAll connections are made via a 21 contact, pluggable screw termina

Strany 12 - NOT RECOMMENDED

8800 Series Trace Moisture Analyzer Instruction ManualiWhen calling your representative for technical support, please have your serial numbers availab

Strany 13 - Needle Valve

Section 3: Instrument13to operate from 15 to 30 VDC 0.5A, with an internal 0.5A fast acting fuse. Please follow the appropriate installation procedure

Strany 14 - 2.4 Mechanical Installation

8800 Series Trace Moisture Analyzer Instruction Manual14or 30VDC, instruments for use in division 2 Hazardous Areas have hermetically sealed relays ra

Strany 15

Section 3: Instrument153.3.2.5 Interfacing to the RS-232 optionThe optional RS-232C interface is provided on the P1 connector. The configuration is 96

Strany 16 - 3. Instrument

8800 Series Trace Moisture Analyzer Instruction Manual163.4.2 Display Conventions1. To display characters with the 7 segment numeric display, the foll

Strany 17

Section 3: Instrument17switches described above.• With a front face plate which has labeled holes placed immediately above the circuit board switches,

Strany 18 - 3.3 Installation

8800 Series Trace Moisture Analyzer Instruction Manual18parts per million by volume computed at the sensor pressure (more about pressure later). The L

Strany 19 - 100 to 250VAC, 50/60Hz, 10VA

Section 3: Instrument193.4.4.2 AlarmsThere are two independent optional alarms, they are named HI and LO alarms. Each alarm can activate a single pole

Strany 20 - Section 3: Instrument

8800 Series Trace Moisture Analyzer Instruction Manual205. Push the MODE button. The display will show:the ‘DEWPOINT’ indicator means that we are chan

Strany 21

Section 3: Instrument21Instrument calibration is recommended in approximately 12 month intervals, and the XTR65W sensor should be re-calibrated after

Strany 22 - 3.4 Operating the Instrument

8800 Series Trace Moisture Analyzer Instruction Manual22Under certain conditions, an over (super) saturated sensor may need to be completely dried out

Strany 23

8800 Series Trace Moisture Analyzer Instruction ManualiiThe manufacturer reserves the right to change or modify the product specification and / or app

Strany 24

Section 3: Instrument23• Make sure the instrument is unlocked.• Press the MODE button a few times until the display shows CAL.• Press the DOWN button

Strany 25

8800 Series Trace Moisture Analyzer Instruction Manual24the calibrations are scheduled at 720 hours interval. Use the UP and DOWN buttons to change th

Strany 26

Section 3: Instrument25sequence of computations is as follows:• The capacitance is converted to percent of full scale (-100°C to +20°C dp), using the

Strany 27

8800 Series Trace Moisture Analyzer Instruction Manual262. Selecting the sensor type: In this mode the user can select the software matching the type

Strany 28

Section 3: Instrument27b. Use the UP or DOWN buttons to modify the cal-adjust value as necessary.c. Press the MODE button to go to the next mode, this

Strany 29

8800 Series Trace Moisture Analyzer Instruction Manual28four times will display “1234”. Pressing and holding the “MODE” button, will display “SET UN/L

Strany 30

Section 3: Instrument29dition ceases to exist, the audio indicator is automatically re-enabled, ready to be activated on next alarm condition. • A mom

Strany 31

8800 Series Trace Moisture Analyzer Instruction Manual30Legend: + denotes a beeping accompaniment to the message./ denotes alternately flashing messag

Strany 32

Section 3: Instrument313.7 MaintenanceThe 8800A and 8800B instrument enclosures may be cleaned when necessary using a damp (moist) cloth. The sensor s

Strany 33

8800 Series Trace Moisture Analyzer Instruction Manual32

Strany 34

8800 Series Trace Moisture Analyzer Instruction ManualiiiExamine the 8800 Series Trace Moisture Analyzer package for damage or mishan-dling. If any da

Strany 35

Glossary33Glossary0/24 mA . . . . . . . . . . . . . . . . range of current for the current loop [0mA=minimum reading, 24mA=maximum reading]4/20 mA .

Strany 36

Manual34used for the circuit [see 4/20]dB . . . . . . . . . . . . . . . . . . . . . see - decibelDB25. . . . . . . . . . . . . . . . . . . a style an

Strany 37

Glossary35impedance . . . . . . . . . . . . . . . the ratio of voltage to current at a specified frequency measured in Ohms, a resistor of the same va

Strany 38 - 3.7 Maintenance

Manual36RG6. . . . . . . . . . . . . . . . . . . . A type of coaxial cable - the letters/numbers refer to the electrical characteristic of the cable

Strany 39

8800 Series Trace Moisture Analyzer Instruction Manual Appendices37Appendix A: Flow Diagram of Operating State User Interface˚C DEWPOINTSETSETHIHI/LOS

Strany 40 - Glossary

8800 Series Trace Moisture Analyzer Instruction Manual Appendices38Flow Diagram Operating State User Interface (continued) Manual & Scheduled Self

Strany 41

8800 Series Trace Moisture Analyzer Instruction Manual Appendices39Appendix B: Flow Diagram of Set-Up State User InterfaceHi/LoSETSETSETLoLoLoHiSETSET

Strany 42

8800 Series Trace Moisture Analyzer Instruction Manual Appendices40Appendix C: Sensor Mechanicalflats for 1 " wrench14/~1.45"14mm x 1.25mm

Strany 43

8800 Series Trace Moisture Analyzer Instruction Manual Appendices41Appendix D: Optional Sensor Fittings

Strany 44 - Power ON initialization

8800 Series Trace Moisture Analyzer Instruction Manual Appendices42Appendix E: 8800 Series Trace Moisture Analyzer Circuit Board DimensionsAppendix E

Strany 45 - Calibration

8800 Series Trace Moisture Analyzer Instruction ManualivWarrantyThis instrument is warranted to be free from defects in workmanship and materials. Lia

Strany 46 - Operating

8800 Series Trace Moisture Analyzer Instruction Manual Appendices43sionsStandard mounting (with clip-on compression brackets)Environmentally sealed mo

Strany 47 - Appendix C: Sensor Mechanical

8800 Series Trace Moisture Analyzer Instruction Manual Appendices44Appendix E (continued): 8800B Series Trace Moisture Analyzer Enclosure Mounting Di

Strany 48

8800 Series Trace Moisture Analyzer Instruction Manual Appendices45Appendix F: Sensor/SpanCheck™ Theory of OperationThe Sensor is constructed as a cap

Strany 49 - SIDE VIEWTOP VIEW

8800 Series Trace Moisture Analyzer Instruction Manual Appendices46Appendix G: Dewpoint Response time AnalysisManufacturers of dewpoint sensors will m

Strany 50 - Standard mounting

8800 Series Trace Moisture Analyzer Instruction Manual Appendices47Of course analyzing on a wet-up step will produce much faster results, because all

Strany 51

8800 Series Trace Moisture Analyzer Instruction Manual Appendices48Appendix H: Sample Gas Filter ConsiderationsSample treatment with pre-filtering may

Strany 52 - 2O3 and the

8800 Series Trace Moisture Analyzer Instruction Manual Appendices49Appendx I: 8800 Series Trace Moisture Analyzer Grounding Considerations1. The senso

Strany 53 - Percent Change

8800 Series Trace Moisture Analyzer Instruction Manual Appendices50Appendix J: Analog Output vs. Dewpoint The current output of the 8800 Series Trace

Strany 54

8800 Series Trace Moisture Analyzer Instruction Manual Appendices51Relationship of 4-20mA output and instrument reading in ppmV or LBSInstrument readi

Strany 55 - Tested Instrument (˚C)

8800 Series Trace Moisture Analyzer Instruction Manual Appendices52Appendix K: RS-232C Interface ProtocolThe instrument uses a simple protocol to comm

Strany 56 - Considerations

8800 Series Trace Moisture Analyzer Instruction ManualvTable of Contents1.0 Overview of the 8800 Series Trace Moisture Analyzer...

Strany 57

8800 Series Trace Moisture Analyzer Instruction Manual Appendices53255 seconds. This feature allows the user to log the measurement on a dumb terminal

Strany 58 - 289.1 LBS = 0˚C = 6063.8 ppmV

8800 Series Trace Moisture Analyzer Instruction Manual Appendices54The printer should have at least 80 character columns, and it is best to use contin

Strany 59

8800 Series Trace Moisture Analyzer Instruction Manual Appendices55Appendix L: Procedure for Exchanging 8800 Series SensorsTo exchange sensors one mus

Strany 60

8800 Series Trace Moisture Analyzer Instruction Manual Appendices56Appendix M: Uncertainty in LBS & ppmV calculationsUncertainty of LBS & ppmV

Strany 61

This page is left Blank intentionally

Strany 62

8/15/02VF

Strany 63 - Measured Dewpoint ˚C

This page is left Blank intentionally

Strany 64

8/15/02VF

Strany 65

58Symbols°C ...17°F ...17AAC power voltag

Strany 66

Index59copper tubing ...6copyright ...icorrosive gases ...

Strany 67

8800 Series Trace Moisture Analyzer Instruction ManualviAppendix D: Optional Sensor Fittings...

Strany 68

60IndexNPL ...1Ooil ...3operating state .

Strany 69

Index61Wwarning labels ... iiiwarranty ...ivwater ...

Strany 70

Teledyne Analytical InstrumentsA Business Unit of Teledyne InstrumentsSENSORS • ANALYZERS • SYSTEMSSCIENTIFIC SOLUTIONS16830 CHESTNUT STREET, CITY OF

Strany 71

Section 1: Introduction11.0 Overview of the 8800 Series Trace Moisture AnalyzerThe 8800 Series Trace Moisture Analyzer is a microprocessor based hygro