The Endosafe -PTS™®Portable Test SystemUSER’S GUIDEVersion 71023 Wappoo Road • Suite 43-B, Charleston, SC 29407Phone: 843.766.7575 • Fax: 843.766.75
Endosafe® - Portable Test System User’s GuideCopyright 2003 Page 7 PIPTS101-10OOppeerraattiinnggtthheeRReeaaddeerrBATTERY INFORMATIONThe PTS Reader
Endosafe® - Portable Test System User’s GuideCopyright 2003 Page 8PIPTS101-10If the batteries are to remain unused for extended periods of time,
Endosafe® - Portable Test System User’s GuideCopyright 2003 Page 9 PIPTS101-10NOTE: Once Erase All Records has been selected and confirmed, all re
Endosafe® - Portable Test System User’s GuideCopyright 2003 Page 10 PIPTS101-10C. EXAMPLE OF KEYPAD OPERATION Setting the Time on the PTS Reader:1.
Endosafe® - Portable Test System User’s GuideCopyright 2003 Page 11 PIPTS101-10D. MENU SHORTCUTS FOR PTS READER KEYPAD The following Menu Shortcuts
Endosafe® - Portable Test System User’s Guide Copyright 2003 Page 12 PIPTS101-10Reader OutputDownloading Output to PC using the PTS Logger Software
Endosafe® - Portable Test System User’s GuideCopyright 2003 Page 13PIPTS101-106. Select Log then View to view data in notepad format. Note: The d
Endosafe® - Portable Test System User’s Guide Copyright 2003 Page 14 PIPTS101-10UUssiinnggEEnnddooSSccaann--VVwwiitthhEEnnddoossaaffee®®--PPTTSSffo
Endosafe® - Portable Test System User’s Guide Copyright 2003 Page 15 PIPTS101-105. Enter the Cartridge lot number. The cartridge lot number will
Endosafe® - Portable Test System User’s Guide Copyright 2003 Page 16 PIPTS101-1013. Upon completion of the assay, a message will be displayed instr
Endosafe® - Portable Test System User’s Guide Copyright 2003 PIPTS101-10TTaabblleeooffCCoo
Endosafe® - Portable Test System User’s Guide Copyright 2003 Page 17 PIPTS101-10TTrroouubblleesshhoooottiinnggListed in the table below are various
Endosafe® - Portable Test System User’s GuideCopyright 2003 Page A1 PIPTS101-10AAppppeennddiixxAAEndosafe® - PTS Endotoxin PERFORMING ROUTINE END
Endosafe® - Portable Test System User’s GuideCopyright 2003 Page A2 PIPTS101-104. Dispense the Sample Once all test information is entered, the P
Endosafe® - Portable Test System User’s Guide Copyright 2003 Page A3 PIPTS101-10RETRIEVING RESULTS OPTIONS1. You can download results directly to
Endosafe® - Portable Test System User’s GuideCopyright 2003 Page B1 PIPTS101-10AAppppeennddiixxBBEndosafe® - PTS Rapid Micro MethodsGram ID PERFO
Endosafe® - Portable Test System User’s Guide Copyright 2003 Page B2PIPTS101-10Note: If refrigerated, allow the cartridge to come to room temper
Endosafe® - Portable Test System User’s Guide Copyright 2003 Page B3 PIPTS101-10GRAM ID TEST RESULTS1. When the test is complete, the PTS Reader
Endosafe® - Portable Test System User’s GuideCopyright 2003 Page B4 PIPTS101-10RETRIEVING RESULTS OPTIONS1. You can download results directly to
Endosafe® - Portable Test System User’s GuideCopyright 2003 Page C1 PIPTS101-10AAppppeennddiixxCCEndosafe® - PTS Inhibition/Enhancement Screen PE
Endosafe® - Portable Test System User’s GuideCopyright 2003 Page C2 PIPTS101-10xEnter Sample 2 Lot # (Sample 2 is the next well to the right of s
Endosafe® - Portable Test System User’s Guide Copyright 2003 PIPTS101-10AAppppeennddiixxee
Endosafe® - Portable Test System User’s Guide Copyright 2003 Page C3 PIPTS101-10INHIBITION/ENHANCEMENT TEST RESULTS1. When the test is complete,
Endosafe® - Portable Test System User’s GuideCopyright 2003 Page C4 PIPTS101-10RETRIEVING RESULTS OPTIONS1. You can download results directly
Endosafe® - Portable Test System User’s GuideCopyright 2003 Page D1 PIPTS101-10AAppppeennddiixxDDEndosafe® - PTS Rapid Micro MethodsGlucanPERFORM
Endosafe® - Portable Test System User’s GuideCopyright 2003 Page D2 PIPTS101-10x Enter Sample ID (Selecting and scrolling with the menu key under
Endosafe® - Portable Test System User’s GuideCopyright 2003 Page D3 PIPTS101-10RETRIEVING RESULTS OPTIONS1. You can download results directly to
Endosafe® - Portable Test System User’s GuideAAppppeennddiixxEEAAbbbbrreevviiaattiioonnssuusseeddiinntthhiissUUsseerr’’ssGGuuiiddee::PPTTSSPPoorrtta
Endosafe® - Portable Test System User’s GuideCopyright 2003 Page F1 PIPTS101-10AAppppeennddiixxFFDDeeccoonnttaammiinnaattiioonnPPrroocceedduurree
Endosafe® - Portable Test System User’s Guide Copyright 2003 Page 1PIPTS101-10SSAAFFEETTYYFFIIRRSSTTWARNINGx DO NOT use an AC power adapter othe
Endosafe® - Portable Test System User’s GuideCopyright 2003 Page 2PIPTS101-10EEnnddoossaaffee®®--PPTTSS––SSyysstteemmOOvveerrvviieewwThe Endosafe
Endosafe® - Portable Test System User’s Guide Copyright 2003 Page 3PIPTS101-10x Dimensions: 9.25” x 4.625” x 2.50” corresponding 235 mm (D) x 12
Endosafe® - Portable Test System User’s Guide Copyright 2003 Page 4PIPTS101-10Endosafe®- PTS Accessory Products ProductDescription CodePC Cable Mo
Endosafe® - Portable Test System User’s GuideCopyright 2003 Page 5 PIPTS101-10Endosafe®- PTS Service ProgramsWARRANTY with PURCHASECharles River La
Endosafe® - Portable Test System User’s Guide Copyright 2003 Page 6 PIPTS101-10Charles River Laboratories will arrange transfer for the PTS instrum
Komentáře k této Příručce