Advanced Force Measurement
Model 9840
Digital Process Monitor
Interface, Inc. 7401 E. Butherus Dr. Scottsdale, Arizona 85260 USA
Tel: (480) 948-5555, Fax: (480) 948-1924
Zobrazit stránku 0
1 2 3 4 5 6 ... 58 59

Shrnutí obsahu

Strany 1 - Digital Process Monitor

Advanced Force Measurement Model 9840 Digital Process Monitor Interface, Inc. 7401 E. Butherus Dr. Scottsdale, Arizona 85260 USA Tel: (480)

Strany 2 - CONTENTS

INTERFACE MODEL 9840 VER 6.0.2 RUN MODE The Model 9840 has three modes of operation, Run mode, Setup mode, and System Calibration mode. At power-on i

Strany 3

INTERFACE MODEL 9840 VER 6.0.2 The table below summarizes the options. Whether the user sees torque or load for each channel depends upon the cell ty

Strany 4

INTERFACE MODEL 9840 VER 6.0.2 The load or torque is always reset to zero. A single channel unit will automatically set Load A or Torque A to zero. R

Strany 5 - INTRODUCTION

INTERFACE MODEL 9840 VER 6.0.2 Print Button If the Print option is purchased, the Print Button can be used to either print data or to freeze the displ

Strany 6

INTERFACE MODEL 9840 VER 6.0.2 SETUP MODE SUMMARY The table below summarizes the setup mode menus. Detailed information about each individual item is

Strany 7 - QUICK START

INTERFACE MODEL 9840 VER 6.0.2 SETUP MODE Setup mode is used to change the setup of the Model 9840. To enter setup mode press the two end buttons at

Strany 8

INTERFACE MODEL 9840 VER 6.0.2 Entering Numerical Data At times you will need to enter numerical data into the Model 9840. The method is consistent i

Strany 9

INTERFACE MODEL 9840 VER 6.0.2 1. Source: -- This entry allows you to scroll through a list (using the left/right or plus/minus buttons) to select t

Strany 10 - RUN MODE

INTERFACE MODEL 9840 VER 6.0.2 2. Full = 4.0000 Lb Zero = 4.0000 Lb (Input) Mv/V (Output)Volt (Output) MA 4 0v 12mA 3 -2.5v 10mA 2 -5.0v 8mA 1

Strany 11 - Tare Button

INTERFACE MODEL 9840 VER 6.0.2 return to the setup mode main menu without changing the cell that is selected. The enter button will select the curren

Strany 12 - Esc Button

INTERFACE MODEL 9840 VER 6.0.2 CONTENTS CONTENTS ...

Strany 13 - Unit Button

INTERFACE MODEL 9840 VER 6.0.2 >>Cal Check Selecting this item will run a calibration check on the cell attached to Channel A or B. The Mod

Strany 14 - SETUP MODE SUMMARY

INTERFACE MODEL 9840 VER 6.0.2 set the torque. When the Model 9840 is reading the cell it will display the “Reading...” message for about 10 seconds.

Strany 15 - SETUP MODE

INTERFACE MODEL 9840 VER 6.0.2 7. Trip if source < / > Set -- Choosing source>set means that the limit will be activated when the source be

Strany 16 - PUB. 2856-16

INTERFACE MODEL 9840 VER 6.0.2 >>Com Address This item allows you to set the address of this unit used for the RS232 ASCII command set. Val

Strany 17

INTERFACE MODEL 9840 VER 6.0.2 >>Decimals Ch A/B This lets you set the maximum number of digits to the right of the decimal that are display

Strany 18 - >Sensor Select

INTERFACE MODEL 9840 VER 6.0.2 REMOTE OPERATION Digital Inputs There are 4 digital inputs available on the Model 9840. They are accessible on the Dig

Strany 19 - >Calibration

INTERFACE MODEL 9840 VER 6.0.2 Communications Settings The Model 9840 supports remote operation using a standard Remote interface consisting of ASCII

Strany 20 - >>Cal Check

INTERFACE MODEL 9840 VER 6.0.2 RS232 RUN MODE COMMANDS The commands listed in this section mirror (and in some cases extend) the functions that are av

Strany 21 - >Limits

INTERFACE MODEL 9840 VER 6.0.2 These are the units for Velocity: 00 – I/M 01 – C/M Front Panel Display Command (F) This command is u

Strany 22 - >>Auto Zero Ch A/B

INTERFACE MODEL 9840 VER 6.0.2 @123F1 Acknowledge: @123 Active Display shows Load A in Lb Other Display shows Peak A in Lb Front Panel Pointer Control

Strany 23

INTERFACE MODEL 9840 VER 6.0.2 Communications Settings ...

Strany 24 - >>Filter Window

INTERFACE MODEL 9840 VER 6.0.2 The repeat number is exactly like the one channel command. It must be 0, 1, or 2. Choosing repeat 1 will make the Mo

Strany 25 - USB Driver

INTERFACE MODEL 9840 VER 6.0.2 @123R1000000 Acknowledge: @123 Reset – Tare A Example 2: @123R0111000 Acknowledge 2: @123 Reset – Peak A Valley A Tar

Strany 26 - Communications Format

INTERFACE MODEL 9840 VER 6.0.2 RS232 SETUP MODE COMMAND SUMMARY @(addr) ╠═ U User Data Entry ║ ╠═ V View current settings ║ ╠═ A

Strany 27 - Question Mark Command (?)

INTERFACE MODEL 9840 VER 6.0.2 RS232 SETUP MODE COMMANDS The commands in this section allow you to change the setup of the Model 9840. You will notic

Strany 28

INTERFACE MODEL 9840 VER 6.0.2 Analog Output Command (A) This command is used to view or set the analog output of the Model 9840. There are two subco

Strany 29

INTERFACE MODEL 9840 VER 6.0.2 Example: @123SA Acknowledge: @123 This is the list of cell calibration data: Ch A = S/N 123456, 100.00 Lb ,

Strany 30 - Print Command (P)

INTERFACE MODEL 9840 VER 6.0.2 Example: @123SD13368# Acknowledge: @123 Deleted Sensor S/N 13368 Ch A = S/N 123456, 100.00 Lb , 3.00150 mV/v, 10.

Strany 31 - Text Display Command (T)

INTERFACE MODEL 9840 VER 6.0.2 command to stop this calibration. Another method of overwriting existing data is to delete the sensor first and then c

Strany 32

INTERFACE MODEL 9840 VER 6.0.2 Calibration Escape Command (CE) The Calibrate Escape (CE) command is used to cancel a calibration that was started with

Strany 33 - User Data Entry Command (U)

INTERFACE MODEL 9840 VER 6.0.2 Calibration by mV/V Mass 1 (CMVM1) command. The mass is entered using the following format: CMVM1(mass value)#. Enteri

Strany 34 - Sensor Command (S)

INTERFACE MODEL 9840 VER 6.0.2 Option View Command (OV)...

Strany 35 - Sensor Delete Command (SD)

INTERFACE MODEL 9840 VER 6.0.2 Calibration by mV/V Torque 1 (CMVT1) command. The torque is entered using the following format: CMVT1(torque value)#.

Strany 36 - Calibration Command (C)

INTERFACE MODEL 9840 VER 6.0.2 Acknowledge: @123 Calibrate by Mass – 5 Point Ready for CMP1 command Calibrate by Masses Point Command (CMP

Strany 37

INTERFACE MODEL 9840 VER 6.0.2 10.00 V , Cal on Oct27-99, 49.532 Lb Shunt unused S/N 13368, 10.00 Lb , 3.00121 mV/v,

Strany 38

INTERFACE MODEL 9840 VER 6.0.2 from the cell before entering this command since it will perform the shunt calibration measurement that is recorded for

Strany 39

INTERFACE MODEL 9840 VER 6.0.2 The Calibration counts per Inch View (CIV) command displays the current setting of the encoder counts per inch constan

Strany 40

INTERFACE MODEL 9840 VER 6.0.2 The limit number chooses which limit (1-4) you want to setup. The normal position should be set to 0 for normally open

Strany 41

INTERFACE MODEL 9840 VER 6.0.2 @123LE Acknowledge: @123 Limit Setup Command Canceled Limit Reset Command (LiR) The Limit Reset (LiR) command is used t

Strany 42

INTERFACE MODEL 9840 VER 6.0.2 Example: Option Printer Command Code Baudrate 4 4800 5 9600 6 19.2K 8 57.6K 9 230.4K @123OP9 Acknowledge: @123 Printer

Strany 43

INTERFACE MODEL 9840 VER 6.0.2 Option Com Baudrate Command (OB) Option Baudrate CommandCode Baudrate 0 300 1 600 2 1200 3 2400 4 4800 5 9600 6 19.

Strany 44 - Limit Set Command (LiS)

INTERFACE MODEL 9840 VER 6.0.2 @123 RS232 EOT is on. Display Options Command (D) The Display options command has 5 subcommands used to set options t

Strany 45 - Limit Setup Escape (LE)

INTERFACE MODEL 9840 VER 6.0.2 INTRODUCTION The Model 9840 is a CE compliant and versatile precision instrument intended for the digital readout of st

Strany 46 - Option View Command (OV)

INTERFACE MODEL 9840 VER 6.0.2 Display Count By Command (DC) The Display Count by (DC) command sets the smallest number that the display will coun

Strany 47

INTERFACE MODEL 9840 VER 6.0.2 Disabling the window sets this value to 0. The window, once enabled and set, will take effect immediately and will rem

Strany 48

INTERFACE MODEL 9840 VER 6.0.2 SYSTEM CALIBRATION MODE The following procedure should be followed to calibrate the Model 9840 itself. This calibratio

Strany 49 - Display Options Command (D)

INTERFACE MODEL 9840 VER 6.0.2 APPENDIX A -- MODEL 9840 SPECIFICATIONS Transducer Interface Excitation 5VDC or 10 VDC software/auto selectable Curr

Strany 50

INTERFACE MODEL 9840 VER 6.0.2 Digital I/O Digital Inputs Interface 4 optically isolated inputs Function remote tare, peak/valley reset, positio

Strany 51

INTERFACE MODEL 9840 VER 6.0.2 Digital Display Display Characters 2 lines by 20 characters Display Update > 4/Sec Scaling Automatic or manual

Strany 52 - SYSTEM CALIBRATION MODE

INTERFACE MODEL 9840 VER 6.0.2 APPENDIX B -- CABLES AND CONNECTORS Load Cell Connectors Notes: DB9 Female Pin Signal 1 EXCITE - HI 2 SENSE - HI1 3

Strany 53

INTERFACE MODEL 9840 VER 6.0.2 Digital I/O Connector Notes: High Density 26 Female Pin Signal 1 LIM 1 A1 2 LIM 1 B1 3 LIM 2 A 4 LIM 2 B 5 LIM

Strany 54

INTERFACE MODEL 9840 VER 6.0.2 USB Connector Pin Name Description 1 VCC +5 VDC 2 D - Data – 3 D+ Data + 4 GND Ground MODEL 9840

Strany 55

INTERFACE MODEL 9840 VER 6.0.2 Precision Shunt Calibration Resistors This diagram shows the connection of the precision internal shunt resistors. T

Strany 56 - Serial Port Connector

INTERFACE MODEL 9840 VER 6.0.2 • Remote operation using standard RS232 ASCII command set. • Each unit is set up to interact with the InterView s

Strany 57 - Chart Rec. Connector

INTERFACE MODEL 9840 VER 6.0.2 QUICK START This section will help you get your Model 9840 set up and operating in just a couple of minutes. The back p

Strany 58 - USB Connector

INTERFACE MODEL 9840 VER 6.0.2 # Message Meaning 1 Interface X.Y.Z This is the version number. 2 Serial # 12345 Option # 114236 This is the seri

Strany 59

INTERFACE MODEL 9840 VER 6.0.2 The rightmost button will change the units that are being used. Repeated presses of this button will cycle through the